EasyManuals Logo
Home>ATC>Chiller>K-Series

ATC K-Series User Manual

ATC K-Series
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
KT1 series chillers Installation and Operation manual
Issue 3.2 5
1.2 Site requirements
Hard, level surface. Ideally smooth to allow freewheeling of castors, which are
designed for indoor use.
Clean, dust free environment. Air-cooled chillers move large volumes of air, and
large amounts of air-borne contamination will result in fouling of the condenser,
reducing the capacity of the unit and in extreme cases causing a system
shut-down.
Non-condensing ambient, from +4°C to +40°C. Cooling capacity is lost above 35°C.
Electrical supply single phase 208VAC +5% (60Hz) 7A, 230VAC +5% (50Hz) 7A,
220VAC +5% (60Hz) 7A.
Electrical terminations.
Live: Brown
Neutral: Blue
Earth: Green/yellow
Clearance front and rear of the unit at least 250mm.
Plumbing to be clean and compatible with the fluid to be used. It is advisable that
the minimum of right-angle bends and compression fittings are used. See also
section 2.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATC K-Series and is the answer not in the manual?

ATC K-Series Specifications

General IconGeneral
BrandATC
ModelK-Series
CategoryChiller
LanguageEnglish