FG_D6_PROFIBUS_EN_01 / 2020-06-17
24 / 68
Results status: (@: 00h 0Fh)
Echo: Acknowledgement of delivery of the master command allowing to determinate in which state
is the slave (current command or command realised).
Error code: In case of command execution error, the corresponding command error bit is activated.
Address (bytes)
Description
00h
Echo
Bit 0 = 1 > Echo reset.
Bit 1 = 1 > Echo start.
Bit 2 = 1 > Echo special cycle.
Bit 3 = 1 > Echo program selection.
Bit 4 = 1 > Echo reading of the results FIFO.
Bit 5 = 1 > Echo reading of the parameters.
Bit 6 = 1 > Echo writing of the parameters.
Bit 7 = 1 > Echo reset of the results FIFO.
01h
Echo
Bit 0 = 1 > Echo reading of the instrument configuration.
Bit 1 = 1 > Echo reading of the configuration / extended menu bits.
Bit 2 = 1 > Echo reading of the function bits.
Bit 3 = 1 > Echo writing of the configuration / extended menu bits.
Bit 4 = 1 > Echo writing of the function bits.
Bit 5 = 1 > Echo reading of the program name.
Bit 6 = 1 > Echo writing of the program name.
Bit 7 = 1 > Echo reading last result.
02h
Error code
(≠ FFh)
Bit 0 = 1 > Reset error.
Bit 1 = 1 > Start error.
Bit 2 = 1 > Special cycle error.
Bit 3 = 1 > Program selection error.
Bit 4 = 1 > Reading of the results FIFO error.
Bit 5 = 1 > Reading of the parameters error.
Bit 6 = 1 > Writing of the parameters error.
Bit 7 = 1 > Reset of the results FIFO error.
03h
Error code
(≠ FFh)
Bit 0 = 1 > Reading of the instrument configuration error.
Bit 1 = 1 > Reading of the configuration bits error.
Bit 2 = 1 > Reading of the function bits error.
Bit 3 = 1 > Writing of the configuration bits error.
Bit 4 = 1 > Writing of the function bits error.
Bit 5 = 1 > Reading of the program name error.
Bit 6 = 1 > Writing of the program name error.
Bit 7 = 1 > Reading last result error.
04h 05h
Reserved.
06h 07h
Current program in use.
08h 09h
Number of results in FIFO (quantity of available results recorded in the FIFO).
0Ah 0Bh
Type of test in progress.