EasyManua.ls Logo

ATEQ F670 - Page 27

ATEQ F670
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QSG_F670_363.00_EN_01_2021-02-09
27/42
Direct mode Sealed part test option
Pressure from 0 to 500 kPa (72.5 PSI).
This configuration can be used for test of small test part volumes.
Protect volumes and pipes from air blowing and temperatures variations.
Connection
Option / description
Air supply to
1
Connection of the air supply to the filter input (0.6 MPa(87 PSI))
Vacuum to
2
Connection to the vacuum (option)
3
to
4
Connection of the reference output to the reference part
5 to 6
Connection of the pressurization output to a transfer volume
6
7
to
9
Connection of the test output to the bell
9
including the part under test
8

Other manuals for ATEQ F670