Timer Max is the maximum time the pressure level can be above
the Run Start level prior to the signal rising above the Clutch
Level. Timer Max is the maximum time for the fastening process.
Time Clutch is the minimum time that the pressure level must be
above the Clutch Level setting at the end of a fastening.
Timer Bump is an ignore timer. In most cases, Timer Bump can
and should be set to zero.
Cycle Start is the pressure level at which the RE Qualifier will
store data.
Run Start is the pressure level which starts Timer Min and Timer
Max. While the pneumatic signal is above Run Start but below
the Clutch Level, the RE Qualifier assumes that the fastener is
fastening.
The Clutch Level is the pressure at which the pneumatic signal is
considered to be clutched out. Both Timer Min and Timer Max
stop running when the pressure reaches this level. Timer Clutch
is activated when the pressure rises above the Clutch Level.
The Batch setting represents the number of fastenings that must
be completed before a Batch OK signal is asserted.
The Total is the number of batches that have been completed.
RE Qualifier System Manual 14