Operating Conditions
The Avive AED™ should only be used in the following
environmental conditions:
32ºF – 122ºF (0ºC - 50ºC)
5 - 95% non-condensing
-280 ft. to 15,000 ft. (-85 m to 4572 m)
The Avive AED is rated to IP54 (with Avive Pad
Cartridge installed) - dust protected, splash and
spray proof.
IMPORTANT: Do not expose the Avive AED to
any liquids or dust when a Pad Cartridge is not
installed.
Temperature
Humidity
Altitude
Water & Ingress
SECTION 6
24
SECTION 6 OPERATING & STORAGE CONDITIONS
Operating & Storage Conditions