EasyManuals Logo

AVL 9180 Service Manual

AVL 9180
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
16
10 Calibration Sequence - 9120, 30, 40
CALIBRATION?
Calibration in
Process 124
Clear Sample
Path
Calibrate
Sample Sensor
Sample Sensor
Voltage ok?
Prime
Standard B
Standard B
detected?
Prime
Standard A
Standard A
detected?
Pressurize Ref.
Asp. Std. A into
Meas. Chamber
Reference
Housing ok?
A
***CHECK***
SAMPLE SENSOR
**STANDARD B**
NOT DETECTED
**STANDARD A**
NOT DETECTED
***CHECK***
REF HOUSING
YES
YES
YES
YES
YES
NO
NO
NO
NO
A
Asp. Std. A into
Meas. Chamber
Standard A
detected?
Obtain Electrode
Voltages Std. A
3 Std. A Values
obtained?
Are Values
reproducible?
Aspir. Std. B into
Meas. Chamber
Standard B
detected?
Obtain Electrode
Voltages Std. B
Perform
Wash Cycle
Aspir. Std. A
into Meas.
Chamber
Standard A
detected?
**STANDARD A**
NOT DETECTED
NO
Calculation of
Std.A-Std.B
(slope)
Slope of all
Electrodes in
Range?
At least 1 slope
in Range?
**STANDARD B**
NOT DETECTED
NO
**STANDARD A**
NOT DETECTED
NO
***CHECK***
ELECTRODES
6 Aspir. of
Std.A performed?
***CHECK***
ELECTRODES
READY 12:45 READY 12:45
XX NOT CAL'D
NO
NO
NO
NO NO
YES YES
YES
YES
YES
YES
YES
YES

Table of Contents

Other manuals for AVL 9180

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AVL 9180 and is the answer not in the manual?

AVL 9180 Specifications

General IconGeneral
BrandAVL
Model9180
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals