10 Calibration Sequence - 9180 (2)
Aspir. Std. A into
Meas. Chamber
Standard A
detected?
**STANDARD A**
NOT DETECTED
NO
Calculation of
Std.A-Std.C
(Use last Std. A)
All Std.A
volt. reprod.?
At least 1 slope
in Range?
***CHECK***
ELECTRODES
NA K LI READY
NA K LI READY
XX, XX NOT CAL'D
NO
YES
YES
B
Aspir. Std. C into
Meas. Chamber
Standard C
detected?
**STANDARD C**
NOT DETECTED
NO
Obtain Electrode
Voltages Std. C
Perform
Wash Cycle
Aspir. Std. A into
Meas. Chamber
Standard A
detected?
**STANDARD A**
NOT DETECTED
NO
Obtain Electrode
Voltages Std. A
Slope A-B and A-C
ok for all Electrodes?
NA K LI READY
XX, XX NOT CAL'D
C
C
YES
YES
YES
YES
NO
NO