Error Messages and Troubleshooting AVL 4000
2-6 Service Manual
2.3 Linearity / Opacity Error,
2.3.1 Opacity Measuring Chamber, Type A
If either of these messages occur, proceed as follows.
Fold the optical elements of the opacity measuring chamber to the side, out of the
way;
Check that both lamps are working - they must both be on.
Replace lamps, if necessary (see section 6.9 Lamp Module).
Clean the glass disc in front of the optical elements.(see Operating Manual, Cleaning
the Optical Elements of the Opacity Measuring Chamber).
Fold the optical elements back into place and acknowledge the message by pressing
>> - you will then see the main opacimeter menu.
2.3.2 Opacity Measuring Chamber, Type 4000
Remove the lens protector from the opacity measuring chamber and check the windows
clean if necessary.
Press << to start an automatic calibration.
If the instrument calibrates itself successfully, you will find yourself in the main
menu.
If the calibration fails, the error message is displayed again. Press >> to call up the
diesel service screen for further error diagnosis.