2 Instruction 3015-4148
HGM300 - Installation Considerations
Locating the Monitor
The HGM300 should be centrally located in the facility and should be easily accessible for visual monitoring
and servicing. Intake sample lines can be up to 500 feet in length, but it is important to remember that
sampling cycle time is proportional to the total number and length of individual sample lines.
Dirt, grease, and oils can adversely affect the operation of the HGM300. The monitor should be installed out
of direct sunlight in a clean, dry area that is not subject to temperature or humidity extremes. Installation of the
monitor in a mechanical room is acceptable provided reasonable environmental conditions exist. If there is a
question, consider installing the unit outside of the mechanical room in a cleaner area of the facility.
WARNING Drilling holes in the HGM300 and
RDM800 enclosure will void the warranty.
Please use knockouts provided for electrical
connections.
WARNING - Do not operate this equipment in the
presence of flammable liquids, vapors or aerosols.
Operation of any electrical instrument in such an
environment constitutes a safety hazard.
WARNING - Always de-energize the power supply
before working inside the instrument.
CAUTION - The HGM300 contains sensitive
electronic components that can be easily damaged.
Be careful not to touch or disturb any of these
components.
Inspection
The HGM300 has been thoroughly inspected and tested prior to shipment from the factory. Nevertheless, it is
recommended that the monitor be re-checked prior to installation. Inspect the outside of the enclosure to
make sure there are no obvious signs of shipping damage. Open the door latches and inspect the interior of
the instrument for loose components that may have become dislodged during shipment. If damage is
discovered, please contact the nearest Bacharach Service Center for assistance.

Table of Contents

Other manuals for Bacharach HGM300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bacharach HGM300 and is the answer not in the manual?

Bacharach HGM300 Specifications

General IconGeneral
BrandBacharach
ModelHGM300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals