Page 45 of 50
Remove the Large Detox Sachet and replace with the new sachet in the service kit.
Figure 29: Large Detox Sachet
Remove the Two Large Catalyst Sachets and replace with the two new sachets in the
service kit.
Figure 30: Large Catalyst Sachets
Replace the ceiling panel and the thumbscrews.
Remove the used sleeves from the front screen and dispose of them.
Figure 31: Gloveport Sleeves
Follow the Deep Cleaning ‘Cleaning’ instructions using the appropriate cleaning material
in the service kit, followed by the ‘Reinstalling workstation components’ section
Fit the new sleeves to the workstation using the new O-rings that are in the service kit.