The optional Auxiliary Switch circuit introduces a second power source to the cabinet. All
power sources must be disconnected before the cabinet is electrically safe. Use proper
Lockout/Tagout procedures.
The cabinet may be provided with an optional lift system rated at 115 VAC, 2A.
(Intermittent Duty, Max 1.2 min. on/12 min off).
Environmental Conditions
The cabinet is designed for use under the following conditions:
• Indoor use
• Altitudes up to 2000 meters
• Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at
40°C.
• Main supply voltage 105 to 125Volts AC
• Transient overvoltages according to Installation Categories (OVERVOLTAGE CATAGORIES) II
per
UL/IEC61010-1, 2
nd
Edition
• Pollution Degree 2
Symbols and Terminology
Protective Earth: Any terminal intended for connection to external protective conductor for
protection against electric shock in case of a fault.
General Caution: Refer to instruction manual for information regarding personnel and
environment protection.