6
ZENITH SERIES
6
RECOMMENDED
INSPECTION PERIODS
ITEM
Inspect
on Receipt
Every
3 Months
Every
6 Months
Maintenance Inspection of All Components at Receipt of Shipment
Make sure all parts/components are included (Please see “Unpacking The Bed”).
X
Check all bed components for obvious damage.
X
Inspect the power supply cord for any cuts and/or damage.
X
Check to see all actuator/motor cords are routed and connected properly to the control box.
X
Mechanical Inspection of Assemblies
Inspect all welds on the sleeping surface, frame, and base assemblies for stress fractures.
X
Inspect all fasteners for wear and looseness.
X
IMPORTANT: Lubricate all pivot points, actuator/motor clevis pins, and control arm clevis
pins as needed. White Lithium Grease is recommended.
X
Mechanical Inspection of Casters & Pedal Locking Mechanism
Check the pedal locking mechanism to make sure it engages and disengages properly.
X
Check the bottom foot pads on the foot end casters, making sure they are clean of debris
and undamaged. Replace if needed.
X
Check all casters to ensure that they roll properly and are unobstructed.
X
Electrical Inspection of Control Box, Pendant, and Staff Control
Check the external power cord that plugs into the control box for any chafing, cuts, or wear.
Replace if damaged.
X
Make sure all attaching hardware is securely tightened.
X
Check all electrical connections for any wear or fractures.
X
Check the external backup battery (if you have one). Replace if needed.
X
Check the pendant cord for any chafing, cuts, or wear.
X
Check all pendant functions - check to make sure each button and associated function work
properly (i.e. head section rises when the Head Up button is activated).
X
Electrical Inspection of Actuators/Motors
Check the actuator/motor cords for any chafing, cuts, or wear.
X
Check the range of movement on all motors to ensure they do not bind in the Full Up or
Full Down positions.