The EASYSPRAY device will continue to emit gas for a brief period after the thumb is removed from the clip/plunger.
This delay helps to avoid clogging of the spray head.
For full prescribing information about Tisseel, contact your local sales representative
Insert 9V battery into the EASYSPRAY
pressure regulator device.
Instructions for Circulating Nurse
|
EasySpray Pressure Regulator
Check the gauge on the EASYSPRAY
device for the appropriate pressure
range of 1.5-2.0 bars (21.5-28.5 psi).
Adjust pressure setting by turning the
pressure control knob.
6
Turn the on/off switch on the front side
of the EASYSPRAY to the ON (I) position.
5
Connect Spray Set filters to EASYSPRAY
device. Connect the blue filter (pressure
line) to the blue female luer connector and
the clear filter (sensor line) to the male luer
connector.
4
Use suitable connection tube to connect
the EASYSPRAY device to medical air
(ranging 3.5 – 7 bar / 50 – 100 psi).
3
Connect EASYSPRAY device to IV pole
or cart rail using the clamps on the back
of the device.
21
Instructions for Scrub Nurse
|
Spray Set
Pass the assembled applicator to the
surgeon for spray application. Pass the
end of the connection tube with the
sterile filters to the circulating nurse.
6
Attach the clip (on the end of the sensor
line) by sliding it into the grooves located
on the top of the syringe plunger.
5
Fit the connection tube of the spray set to
the luer-lock connector on the underside of
the spray head.
4
Fasten the pull strap to the double
syringe system to assure the spray head
is tightly secured.
3
Firmly attach the spray head to the
nozzle of the syringes.
2
Prepare TISSEEL Fibrin Sealant according
to the instructions in the package insert.
1
Instructions for Surgeon
Spray from a distance of 10 – 15 cm for
optimum results.
To activate the flow of gas occlude the
opening in the clip center with thumb.
To begin application, gently depress the
syringe plunger.
32
Caution: Any application of pressurized gas may be associated with a potential risk of air or gas embolism, tissue rupture or air or gas entrapment with
compression, which may be life threatening. Be sure to take appropriate measures to address these risks by observing the recommended minimum
spraying distance and the maximum pressure provided in the appropriate spray set instructions for use.
The use of TISSEEL Fibrin Sealant is restricted to experienced surgeons who have been trained in the use of TISSEEL.
In order to ensure optimal safe use of TISSEEL by spray application, apply a minimum spray distance of 10 cm and a maximum spray
pressure of 2.0 bar (28.5 psi) to minimize the potential risk of air or gas embolism, tissue rupture, or air or gas entrapment with
compression.
Confirm (verbally) the actual pressure
with OR personel.
1
EASYSPRAY Quick Reference Guide (for open wound surgery)

Other manuals for baxter EASYSPRAY

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter EASYSPRAY and is the answer not in the manual?

baxter EASYSPRAY Specifications

General IconGeneral
Brandbaxter
ModelEASYSPRAY
CategoryMedical Equipment
LanguageEnglish

Related product manuals