Position DUPLOSPRAY System so the foot pedal is placed next to the surgeon's foot.
Attach gas supply hose to the source of medical grade CO₂.
Connect the blue vent line filter and clear gas line filter to the regulator.
Depress foot pedal, adjust gas flow rate to 1.0-2.0 L/min, check flow gauge.
Thread sterile replaceable tip onto applicator using the tip alignment tool.
Firmly attach the applicator to the nozzles of the double syringe.
Fasten the pull strap to the double syringe holder to assure the spray applicator is tightly secured.
Attach gas supply line to sterile applicator and turn white locking collar to secure.
Pass assembled double syringe and spray applicator to surgeon for application.
Attach patient vent line to trocar vent valve, ensure vent valve is fully open.
Confirm actual flow of max 2 litres/min with theatre personnel.
Depress foot pedal to start gas flow before applying TISSEEL. Check gas flow gauge.
Dispense TISSEEL by depressing syringe plungers slowly. Hold pedal 3-5s after to clear tip.
Advice on vent line use, clearing occluded tips, and twin-tube assembly separation.
Warning about potential risks of air/gas embolism, tissue rupture with pressurised gas.
Recommendations for pressure, distance, and devices for TISSEEL spray application.
Summary of TISSEEL indications, dosage, side effects, precautions, and contraindications.