10. CALIBRATION (ONLY AUTHORIZED PERSONAL)
The calibraon can be only done by the authorized technical personnel, please, if you wish do a calibraon
contact with your authorized calibraon organism who is authorized to seal and to unseal the scale.
In order to enter the calibraon parameter it is necessary to short the K2 jumper K2 of the main keyboard.
Turn the power o.
Turn the power back on, during the counng from 9 to 0 press the key.
The display will show Pn (ask password), press
to enter.
The display wil show the rst funcon, PO CHk. Press unl display show P2 CAL, press to enter,
press to select CAL, press key to enter calibrate. The display will show UnLd.
Remove any weight from the plaorm. Aer stable indicator on, press the key.
Then the display will show the last calibraon weight used. If this is correct you can connue by pressing
the key. If it is not correct use the keys to change the calibraon weight value. When it
is correct press the key.
Then display will show LoAd. Place the calibraon weight on the scale. Aer stable press the key.
If the calibraon is acceptable the display will return to normal. If an error message is shown try calibraon
again as a disturbance may have prevented a successful calibraon.
If the problem persists then contact your dealer.
Aer calibraon, itshould be checked to verify the calibraon and linearity is correct. If necessary repeat
calibraon, especially be certain the scale is stable before accepng any weight.
11. ERROR CODE
ERROR
CODES
DESCRIPTION RESOLUTION
- - - - - Over range
Remove weight from the indicator.
If the problem persist contact your dealer for assistance.
Err4
Zero Seng Error
The indicator was outside the normal zero seng range either when it was
turned on or when the key was pressed.
Remove weight from the scale and try again.
Use the key to set the display to zero value.
If the problem persist contact your dealer for assitance.
Err6
A/D out of range
The values from the A/D converter are outside the normal range. Remove
weight from the scale if overloaded, make sure the pan is aached.
Indicates the load cell or the electronics may be faulty.
If the problem persist contact your dealer.
12. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the de-
livery date.
During this period, GIROPÈS SL will take charge of repairing the scale.
This warranty does not cover defects or damaged caused by misuse, overloading or improper installaon
contrary to the recommendaons described in this manual.
This warranty does not cover shipping costs for the reparaon of the balance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Baxtran BR20 and is the answer not in the manual?

Baxtran BR20 Specifications

General IconGeneral
BrandBaxtran
ModelBR20
CategoryAccessories
LanguageEnglish

Related product manuals