ENUSER MANUAL XTA
7. PARAMETER
7.1 PARAMETER SETTINGS
ENTER INTO THE MENU
Turn on the scales. Press and hold
key during the self test.
Display will show "Pin” briefly PIN is activated. Press the Pin number to get into the parameter menu.
The default PIN number is “0000”.
Press numeric key 0 four times, display will be shown Pn - - - -
• Press
key to confirm, it will enter into the parameter settings and will to show “FO CAL
ENTER TO SELECTED MENU
Press
, for to confirm and access to displayed parameter.
SELECT THE MENU
• Press
or for to choose menu.
RETURN TO WEIGHING MODE
• Press
for to escape from the menu and exit to weighing mode
ACTIVE PARAMETER
!
Press CAL switch to enable metrological parameters for to access and change them. CAL switch is locating bottom
of the scale.
You need to break metrological seals for to press cal switch. With this action you will loose scale verification.
CAL SWITCH