Chapter
5:
System
Testing
2.
With
the
(-)
of
the
DMM
connected
to
TP5,
verify:
TP3
1100+
100
mV,
depending
on
temperature
(values
assume
room
temperature)
Τρ4
53$
+
50
шУ
TP6
315
+
100
mV,
depending
on
temperature
(values.
assume
room
temperature)
3.
With
the
(-)
of
the
DMM
connected
to
J11
pin
2,
verify:
Hipini
77+02V
Jllpin3
-7.7+0.2V
LOX
Board
Power
Supply
and
Reference
A.
Verify
power
supply
operation
by
checking
the
following
voltages:
1.
With
the
(-)
of
the
DMM
connected
to
TPI,
verify:
TP2
50+0.25V
TP3
-5.0+0.25
V
TP10
3.40
V (3.10
to
3.71
V)
Pneumatics
Inlet
Sample
Flow
A.
Connect
the
flowmeter
to
the
moisture
trap
inlet
with
the
8
feet
sample
line.
B.
Verify
a
flow
rate
of
150
(+20/-10)
ml/min.
Occlusion
Verification
A.
Occlude
sample
inlet.
B.
Verify
that
while
in
occlusion
in
about
30
seconds.
C.
Verify
that
while
in
occlusion
pump
turns
on
to
check
if
units
is
still
occluded
every
5
seconds
until
the
occlusion
is
removed.
5-2
4
Capnocheck
Plus
Service
Manual