MA0120(11) (2022-04) • English 9
E35 Incorrect drawer
Tube just scanned belongs in another drawer. Open next drawer and scan tube barcode. Repeat until error
does not recur.
E37 Barcode scan order incorrect
System expected a barcode sequence number and an accession barcode was scanned (or vice versa). Be sure
to scan the type of barcode the system expects.
E38 No storage media in drive Insert blank, formatted, write-enabled storage media in the storage media drive to save data to storage media.
E39 Storage media full, insert another
Insert blank, formatted, write-enabled storage media in the storage media drive to complete the save data to
storage media function.
E40 Network error Start BD EpiCenter™ system.
E43
LIS parameter changed:
instrument must be restarted.
Select OK soft key to implement changes or select not OK soft key to cancel changes.
E45 Drawer error A drawer is not fully closed. Close drawer.
E50 Internal software error System encountered a software error. Save data to storage media and call BD.
E92 AST Set Error Main error code for any station error in an AST set. Refer to User’s Manual for additional information.
Code
Meaning
Corrective Action
(If the recommended actions do not correct the situation, refer to the User’s Manual or contact BD)
Error Codes E35 – E92