What to do if my Beckman Coulter Allegra 25R Laboratory Equipment experiences Chamber overtemperature?
- DDiana LoveAug 17, 2025
If your Beckman Coulter Laboratory Equipment indicates a chamber overtemperature, follow these steps: 1. Turn the power off. 2. Verify that the ambient temperature is within the specified range of 10 to 35°C. 3. Inspect the air inlet and exhaust vents for any obstructions and clean them if necessary. 4. Allow the motor to cool down for about 15 minutes with the door open, then turn the power back on to reset the thermal protector circuit.