What to do if Beckman Coulter Allegra 6 Laboratory Equipment rotor cannot achieve set speed?
- SsherriwhitakerAug 15, 2025
If the rotor speed of your Beckman Coulter Laboratory Equipment cannot reach the set speed, there are several potential causes and solutions: * **Low line voltage:** Measure the line voltage while the centrifuge is operating. * **Imbalance detector tripped:** Turn the POWER switch to OFF (O), wait 30 seconds, and then turn it back ON (I). * **Thermal protector tripped:** Turn the POWER switch to OFF (O), wait 60 seconds, and then turn it back ON (I). * **Electrical failure:** Check the connections. If the issue persists, further assistance is needed. * **Worn motor components or motor failure:** If the motor bearings, commutator are worn, or the motor has failed, further assistance is needed.