What to do for imbalance error on Beckman Coulter Allegra X-14?
- LLaura LeeAug 21, 2025
If you receive an imbalance error on your Beckman Coulter Laboratory Equipment, there are several potential causes: * The rotor load might be severely out of balance. Ensure that tubes or bottles are balanced and loaded symmetrically in the rotor as described in the applicable rotor manual. * The rotor tie-down might not be properly fastened. Make sure the rotor lid is properly aligned and the tie-down T-bar is securely fastened. * The centrifuge might be misaligned (tilted). Level the centrifuge on the bench or table. * The centrifuge might have been moved during operation. After the rotor comes to a complete stop, open and close the centrifuge door, then restart.