How to fix door latch on Beckman Coulter Laboratory Equipment?
- SSteven PetersonAug 19, 2025
If the latches are not operating properly, press down on both sides of the door until the DOOR LED comes on, then press CE to clear the message.
How to fix door latch on Beckman Coulter Laboratory Equipment?
If the latches are not operating properly, press down on both sides of the door until the DOOR LED comes on, then press CE to clear the message.
What to do if I can't open the door on Beckman Coulter Laboratory Equipment?
If the door latches do not release, press down on the door, then press DOOR again. If the door opens but the message stays on the display, press CE to clear it.
What to do if Beckman Coulter Laboratory Equipment has a power failure?
If a power failure occurred due to a system or facility power failure, press CE to clear the message and restart the run. The remaining time will show on the display.
What to do if the rotor temperature exceeds 4C above set on Beckman Coulter Laboratory Equipment?
If the rotor temperature exceeds the temperature setting by more than 4°C but less than 8°C after equilibration, press CE to clear the message.
What to do if Beckman Coulter Laboratory Equipment shows 'no rotor match'?
If the system cannot identify the rotor, make sure the rotor in use is a compatible Beckman Coulter rotor and that magnets are in place on the rotor bottom. Press CE to clear the message.
What to do if Beckman Coulter Laboratory Equipment shows a drive error?
If there is a drive system problem, after the run stops, press CE to clear the message, then restart the run.
How to fix rotor ID problem on Beckman Coulter Laboratory Equipment?
If the rotor ID has a problem because no magnets are identified, or magnets are incorrectly identified, check that magnets are in place on the rotor bottom and press CE to clear the message.
What to do if Beckman Coulter Laboratory Equipment shows a rotor ID mismatch?
If the entered rotor number is not the same as the rotor identified, press CE to clear the message and enter the correct rotor entry identification.
What to do if Beckman Coulter Laboratory Equipment shows an imbalance error?
If there is a system imbalance error, press CE to clear the message, then restart the run.
Why chamber does not reach selected temperature in Beckman Coulter Laboratory Equipment?
If the centrifuge cannot maintain the selected temperature for the rotor in use at the selected speed, precool or prewarm rotors before running at low or high temperatures. If the air intake vent is blocked, make sure the air intake vent is clear.
Brand | Beckman Coulter |
---|---|
Model | Allegra X-12 Series |
Category | Laboratory Equipment |
Language | English |
Explains hazard symbols and their meanings for safe operation.
Covers safety aspects of installing and maintaining the centrifuge.
Details electrical safety precautions for operating the centrifuge.
Provides guidelines for the safe mechanical operation of the centrifuge.
Addresses safety precautions for handling chemical and biological materials.
Explains China RoHS compliance labels and requirements for electronic products.
States manufacturing standards and compliance with regulatory agencies.
Outlines the manual's purpose and content organization.
Explains formatting conventions used in the manual for clarity.
Explains the WEEE directive symbol and disposal guidelines for electronic equipment.
Details the primary functions and built-in safety features of the centrifuge.
Describes the physical construction and components of the centrifuge.
Introduces the main controls and indicators on the centrifuge.
Lists technical specifications of the Allegra X-12 Series centrifuges.
Lists compatible rotors with their specifications and codes.
Provides an overview of preinstallation requirements.
Details the necessary space, ventilation, and clearances for installation.
Specifies power cord details and grounding requirements.
Explains the need and method for securing the centrifuge.
Recommends performing a test run after installation.
Provides an overview of centrifuge operating procedures.
Step-by-step guide for performing a manual centrifuge run.
Step-by-step guide for executing a programmed centrifuge run.
Explains how to select and confirm the installed rotor.
Guides on setting the desired rotor speed in RPM or RCF.
Explains how to set the run duration or use hold mode.
Describes how to select ramp rates for acceleration and deceleration.
Outlines the steps to initiate a centrifuge run.
Explains the use of the pulse key for short-duration centrifugation.
Instructions on how to control audible beeps for keys, errors, and run completion.
Overview of troubleshooting procedures and contacting service.
Explains how to interpret and clear user messages and diagnostic codes.
Lists common diagnostic messages, problems, results, and recommended actions.
Addresses issues not indicated by diagnostic messages.
Explains manual procedure for sample retrieval during power outages.
Details how to manually override the door latch for sample access.
General advice for maintaining the centrifuge's condition.
Lists essential maintenance tasks for continued performance.
Step-by-step guide for cleaning the centrifuge and its components.
Outlines steps for decontaminating after exposure to hazardous materials.
Provides guidance on storing and transporting the centrifuge safely.
Lists available parts and supplies for ordering.
Details the warranty terms, conditions, and disclaimers for the centrifuge.