English ⋅ Nautilus
®
CC plus ⋅ Nautilus
®
T
75
86105 BA-en/00
Checking the seals
You can check the tightness of the seal after the O-
rings have been cleaned.
During the test the chambers are filled with
compressed air. If no air escapes (generation of
noise!), the O-rings function properly.
1. Go to Settings > O-rings > Check O-rings, see
page 74.
2. If air escapes, the O-ring need to be cleaned.
Draining the pressure reducers
The pressure reducer must be drained every 500
castings.
1. Switch the device off and pull out the mains
plug.
2. Remove the cover (1) on the left side (loosen
the slotted screws).
3. Have a container or a cloth at hand to collect
the water.
4. Turn the bottom screw (2) counterclockwise
until it stops to drain the water.
Reactivating fuses
Circuit-breakers against overcurrent are located on
the rear side of the device (see circuit diagram).
1. Wait for approx. 1 minute after the device has
been switched off.
2. Reactivate fuses that have been tripped by
pushing them back in place.
3. If fuses are triggered repeatedly, notify Cus-
tomer Service.