English ⋅ Nautilus
®
CC plus ⋅ Nautilus
®
T
79
86105 BA-en/00
Troubleshooting
The NAUTILUS
®
-devices issue three types of
messages.
• Error messages indicate significant errors,
which will abort the casting process.
• Warnings will not abort the casting process;
after clearing the message (OK button), the
process can be continued. Exception: W001
(insufficient cooling water) and W002 (insuffi-
cient compressed air).
• Notes refer to general service and mainte-
nance tasks. They are displayed before or after
casting.
General notes on troubleshooting:
1. Switch the device off (exceptions: see *).
2. Eliminate the malfunction.
3. Switch the device on.
4. Notify Service if messages appear repeatedly.
Service must only be carried out by authorized
BEGO repair shops!
Error messages
Malfunction Cause Remedy
Main switch is on, display is
blank
Device has no power. Check power supply connection;
Check fuses (→ p. 75).
There is a hissing sound during
operation.
Sealing surfaces on crucible
and/or mould door are soiled/
dirty.
Caution! Components are very hot!
Clean seals and sealing surfaces
(→ p. 74).
No vacuum and/or compressed
air.
Air filter is dirty. Replace the air filter (→ p. 77).
Error message Cause Remedy
E001 No water Add water (→ page 72).
E002 No compressed air * Ensure compressed air supply ≥ 5 bar (100 l/min) or
switch off other compressed air loads.
Install compressed air reservoir, as needed
(→ page 16).
Clean seals (→ page 74).
E003 Lock not activated * Close doors properly.
E010 Generator malfunction Switch the device off!
Check fuses (→ p. 75).
Check induction coil (→ p. 70): Remove any dirt that
has collected between the coil windings
Notify Customer Service.