Installation
Cooling water
Adding cooling water (see page 72).
Compressed air
1/4 "
5 .... 8 bar (0.5 .... 0.8 MPa) (72.5 .... 116 psi)
100 l / min
Insufficient supply of compressed air during
pressure casting (warning message W011) can be
compensated with the compressed air reservoir
(REF 16260).
Insufficient supply of compressed air during evacu-
ation (warning message W002, W013) cannot be
compensated with the compressed air reservoir.
Important! Remove protective cap!
WARNING
The warning signs provided (symbol:
"pacemaker") are to be placed at the
entrances to the room in which the
ready-to-use device is located!
NOTICE
After initial commissioning, run the device for at
least 2 minutes to allow for the cooling water to
circulate and check the water level again.