English Nautilus
®
CC plus Nautilus
®
T
81
86105 BA-en/00
en
Note
W 005 Internal battery empty Notify Customer Service.
W 010 Chamber leak Clean seals ( page 74).
W 011 Pressure increase too slow Check compressed air supply 5 bar (100 l/min).
Install compressed air reservoir, as needed
( page 16).
Clean seals ( page 74).
W 012 Pressure increase too fast * Notify Customer Service.
W 013 Maximum vacuum was not
reached
Check compressed air supply 5 bar (100 l/min).
Clean the ejector nozzle ( p. 76).
Replace the air filter ( p. 77).
W 014 Maximum mould pressure
was not reached
Check compressed air supply 5 bar (100 l/min).
Install compressed air reservoir, as needed
( page 16).
Clean seals ( p. 74).
W 015 Critical cooling water tem-
perature
* Allow cooling water to cool down while the device is on.
W 016 Chamber not tight Clean seals ( page 74). If the service message still
persists, notify Customer Service.
Error message Cause Remedy
Error message Cause Remedy
H 001 Initialization of the internal
USB flash drive
H 077 (every
8,000 castings)
Service required Notify Customer Service.
H 099 (after
10,000 castings)
Service required Notify Customer Service.