12/ 2007
BENNING MM P3
25
7.6 Frequency ranges
Overload protection for frequency measurements: 600 V
eff
7.6.1 Frequency ranges for square wave signals (Switch setting: Hz, %)
Measuring
range
Resolution
Measuring accuracy for 5 V
SS
max.
(square wave signal)
Sensitivity
5 Hz 0.001 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
50 Hz 0.01 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
500 Hz 0.1 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
5 kHz 1 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
50 kHz 10 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
500 kHz 100 Hz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
5 MHz 1 kHz ± (0.3 % of the measured value + 5 digits) > 1.0 V
ss
(rectangle)
7.6.2 Frequency indication for sinusoidal signals (Switch setting: V
AC
, Hz, %) and
actuation of the „SELECT“ key:
Measuring accuracy: ± (0.3 % + 5 digits) applies to sinusoidal voltages of up
to 600 V
eff
(10 Hz - 500 Hz) and displayed values in the AC voltage range
(V
AC
) higher than 50 % of the final measuring range value
7.7 Pulse duty ratio for square wave signals (Switch setting: Hz, %)
Overload protection in the case of pulse ratio measurement: 600 V
eff
Meas. range Resolution
Measuring accuracy up to 5 V
SS
(square wave signal, 5 Hz - 5 kHz)
Sensitivity
(30 % ≤ % ≤ 70 %)
0.1 % - 99.9 % 0.1 % ± (0.5 % of the measured value + 3 digits) > 1.0 V
ss
(rectangle)
8. Measuring by means of the BENNING MM P3
8.1 Preparing the measurement
Operate and store the BENNING MM P3 at the specied storage and operating
temperatures only!
Do not permanently expose the device to sunlight.
- The connected black and red safety measuring lines comply with the valid
provision, if they are undamaged.
- Please protect the connected black and red safety measuring lines against
contamination and damages!
- Check insulation of the safety measuring lines. If the insulation is damaged,
multimeter must be replaced immediately.
- Check the safety measuring lines for continuity. If the conductor in the safety
measuring line is interrupted, replace the multimeter immediately.
- Before selecting another function by means of the rotary switch 7, disconnect
the safety measuring lines from the measuring point.
-
Strong sources of interference in the vicinity of the BENNING MM P3
might involve
unstable readings and measuring errors.