Berchtold Operon D760 - Guidelines and Manufacturers Declaration - Electromagnetic Interference Immunity

Default Icon
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurement of
emissions
Compliance Electromagnetic environment
guidelines
Voltage fluctuations
to a public power grid that also
supplies buildings used for residen-
tial purposes.
Fulfilled
11.3 Guidelines and manufacturer's declaration –
Electromagnetic interference immunity
Interference im-
munity checks
IEC 60601 test/compliance
level
Electromagnetic envi-
ronment – guidelines
Electrostatic dis-
charge (ESD) per
IEC 61000-4-2
± 4kV
Contact discharge
Synthetic floors should
be antistatic and the
relative humidity at
least 30%.
±8kV
Air discharge
Electrical fast transi-
ent bursts per IEC
61000-4-4
±2 kV
For power grids
The supply voltage
quality should corre-
spond to that of a typi-
cal commercial or hos-
pital environment.
Surges per IEC
61000-4-5
± 2 kV Countervoltage The supply voltage
quality should corre-
spond to that of a typi-
cal commercial or hos-
pital environment.
± 1 kV Asymmetrical volt-
age
Voltage drops, short
interruptions and
fluctuations in the
power supply volt-
age per IEC
61000-4-11 (V
T
AC
line power voltage
prior to use of test
level)
>5% VT
t=10 ms
(> 95% dip)
The supply voltage
quality should corre-
spond to that of a typi-
cal commercial or hos-
pital environment.
To ensure uninterrupted
operation of the OPER-
ON® OR tables, they
must also be connect-
ed to an emergency
power supply in ac-
cordance with DIN VDE
0100–710.
40% VT
t=100 ms
(60% dip)
70% VT
t=500 ms
(30% dip)
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Appendix
105