Peak Hold button
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture
negative or positive peaks as fast as 1 millisecond in duration.
Turn the function switch to the A or V position.
Use the
MODE button to select AC or DC.
Allow time for the display to stabilize.
Press and Hold the
PEAK button until “CALappears in the display. This procedure will zero
the range selected.
Press the
PEAK button, Pmax will display.
The display will update each time a higher positive peak occurs.
Press the
PEAK button again, Pmin will display. The display will now update and indicate the
lowest negative peak.
To return to normal operation, press and hold the
PEAK button until the Pmin or Pmax indi-
cator switches off.
Note:
If the Function switch position is changed after a calibration the Peak Hold calibration must be
repeated for the new function selected.