Measure-
ment range
Cuff pressure 300 mmHg,
systolic pressure 50 280 mmHg,
diastolic pressure 30 200 mmHg,
pulse 40 – 199 beats/minute
Display
accuracy
Systolic pressure ± 3 mmHg,
diastolic pressure ± 3 mmHg,
pulse ± 5% of the displayed value
Meas-
urement
uncertainty
Max. permissible standard deviation according
to clinical testing: systolic pressure 8 mmHg,
diastolic pressure 8 mmHg
Memory 2 x 100 memory spaces
Dimensions L 140 mm x W 94 mm x H 46 mm
Weight Approx. 437 g (without batteries, with cuff)
Cuff size 22 to 42 cm upper arm circumference
Operating
conditions
+ 10 °C to + 40 °C, 10 – 85 % relative humidity
(non-condensing), 700 1060 hPa ambient
pressure
Storage and
transport
conditions
-20 °C to + 55 °C, ≤ 90% relative humidity
Power
supply
4 x 1.5 V
AAA batteries
Battery life Lasts for approx. 300 measurements de-
pending on the blood pressure and inflation
pressure
Product life
cycle to be
expected
Information on the life cycle of the product can
be found at beurer.com
Classifica-
tion
Internal power supply, IP20 no AP or APG,
continuous operation, type BF applied part
The serial number is located on the device or in the battery
compartment.
Technical specifications are subject to change without notifi-
cation to allow for updates.
This device conforms with the European stand-
ard EN 60601-1-2 (in accordance with CISPR-11,
IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2,
IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5,
IEC 61000-4-6, IEC 61000-4-7, IEC 61000-4-8,
IEC
61000-4-11) and is subject to particular precautions
with regard to electromagnetic compatibility. Please
note that portable and mobile HF communication sys-
tems may interfere with this device.
The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long
useful life. If the device is used for commercial medical
purposes, it must be regularly tested for accuracy by ap-
propriate means. Precise instructions for checking accu-
racy may be requested from the service address.
Mains part
Model no. LXCP12X-050100BG
Input 100 – 240 V, 50 – 60 Hz, 0.5 A max