EasyManua.ls Logo

Beurer BM 85 - Device Care and Troubleshooting; Cleaning and Storage Guidelines; Interpreting Error Messages

Beurer BM 85
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
If your smartphone has a protective cover, remove this to en-
sure that there is no interference during the transfer. Begin the
data transfer in the “beurer HealthManager”-app.
9. Cleaning and storing the device and cuff
Clean the device and cuff carefully using a slightly damp
cloth only.
Do not use any cleaning agents or solvents.
Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.
10. Error messages/trouble-shooting
In case of faults, the _ message appears in the display.
Error messages may appear if:
systolic or diastolic pressure could not be measured ( or
appears on the display)
systolic or diastolic pressure was outside the measurement
range ( or Lo appears on the display)
the cuff is fastened too tightly or loosely ( or appears
on the display)
the pump pressure is higher than 300 mmHg ( appears
on the display)
pumping up takes longer than 160 seconds ( appears on
the display)
there is a system or device error ( , , or ap-
pears on the display)
the data could not be sent to the PC ( appears in the
display).
the data could not be sent via Bluetooth
®
(
).
In such cases, repeat the measurement and/or the data trans-
fer. Ensure that the cuff line is correctly attached and that you
do not move or speak during the measurement.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical specifica-
tions”, the technical alarm will appear on the display indicating
either
  ” or “ Lo . In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.

Table of Contents

Related product manuals