8. Use the left and right ˄/˅setting buttons for and to select an impulse intensity.
EMS 35
Various impulse sequences run in this burst program. Burst programs are suitable for all areas of application to be treated
with changing signal patterns (to minimize the level of accustoming to the treatment). In this program, you can set the impulse
frequency between 1 and 150 Hz and the impulse width between 80 and 450 μs. You can also set the working time and pause
time for this program between 1 and 30 seconds each.
1. Place the electrodes on the desired area for treatment (for placement suggestions see the ‘Electrode Positioning’ section) and
connect them to the device.
2. Select the EMS 35 program as described in the ‘Starting Application’ section, steps 3-5.
3. Use the ˄/˅ setting buttons to select the working time (‘on time’) and use the ENTER button to confirm.
4. Use the ˄/˅ setting buttons to select the pause time (‘off time’) and use the ENTER button to confirm.
5. Use the ˄/˅ setting buttons to select an impulse frequency and use the ENTER button to confirm.
6. Use the ˄/˅ setting buttons to select an impulse width and use the ENTER button to confirm.
7. Use the ˄/˅ setting buttons to select the treatment time and use the ENTER button to confirm.
8. Use the left and right ˄/˅ setting buttons for and to select an impulse intensity.
11. Doctor‘s Function
The Doctor’s Function is a special setting to allow you to access your personal program even more easily and directly. Your
individual program settings are instantly recalled and activated when the device is switched on. You may wish to adjust this
individual program following advice from your doctor.
Setting the Doctor’s Function
1.
S
elect your program and the corresponding settings as described in the ‘Starting Application’ section.
2.
A
t the start of the stimulation treatment, the impulse intensity of and is set to 00 by default. No impulses are sent to the
electrodes yet. Before setting the desired impulse intensity using the intensity setting buttons, press and hold the ˅ button
for five seconds. Storage in the Doctor’s function is confirmed with a long beep.
3.
I
f you switch on the device again, the program you saved using the Doctor’s function opens directly.