EasyManua.ls Logo

Binder VD 115 - User Manual

Default Icon
173 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Issue 03/2020 Art. No. 7001-0383
Operating Manual
Translation of the original operating manual
VD (E3.1)
Vacuum Drying Oven
Standard: with RD4 microprocessor program controller
Optional: with MB2 microprocessor program controller
Model Model version Art. No.
VD 23 VD023-230V 9630-0001
VD 23-UL VD023UL-120V 9630-0005
VD 56 VD056-230V 9630-0002
VD 56-UL VD056UL-120V 9630-0006
VD 115 VD115-230V 9630-0003
VD 115-UL VD115UL-120V 9630-0007
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder VD 115

Questions and Answers

  • S
    scollierAug 21, 2025
    What to do if my Binder VD 115 Oven chamber doesn’t heat up and shows a “Safety controller” alarm?
    • J
      Jason GarciaAug 21, 2025
      If the alarm message “Safety controller” appears on the controller display of your Binder oven, let the chamber cool down. Then, acknowledge the alarm on the controller. Check the settings of the temperature setpoint and safety controller value. If needed, select a suitable safety controller value.
  • E
    Evan SanchezSep 3, 2025
    Why are there deviations from the temperature set-point in the equilibrated state of my Binder oven?
    • M
      Maria AndersonSep 3, 2025
      If there are deviations from the temperature set-point in the equilibrated state of your Binder oven, use only the delivered expansion racks and do not change between aluminum and stainless steel racks.
  • L
    Lauren JohnsonSep 1, 2025
    Why are there deviations from the indicated heating-up times in my Binder VD 115 Oven?
    • N
      Nicole BarkerSep 1, 2025
      If you notice deviations from the indicated heating-up times in your Binder oven, reduce the load in the oven or allow for longer heating-up times.
  • M
    Melanie GrimesAug 30, 2025
    What to do if the controller on my Binder oven does not equilibrate to entered setpoints?
    • J
      John KingAug 30, 2025
      If the controller of your Binder oven does not equilibrate to entered setpoints, turn on temperature control. Turn on pressure control. If you are using MB2 controller, change to Fixed value operation mode if set-points were entered in Fixed value operation mode but the controller is in program operation mode.
  • V
    vcastroAug 29, 2025
    What causes a Binder oven to not hold vacuum?
    • C
      Catherine FisherAug 29, 2025
      If the vacuum is not being held in your Binder oven, replace the door gasket, the safety glass panel, or the gaskets of the small flange connections.
  • B
    boyerwesleyAug 27, 2025
    Why is the chamber of my Binder VD 115 permanently turned off?
    • M
      Mark RandolphAug 27, 2025
      If the chamber of your Binder oven is permanently turned off, check the connection to the power supply. Ensure the power supply matches the voltage requirements of 115V or 230V. Also, check the chamber fuse.
  • A
    ahallAug 27, 2025
    How to fix a Binder VD 115 Oven when the chamber is heating permanently but the set-point is not maintained?
    • C
      Carol DunlapAug 27, 2025
      If your Binder oven chamber is heating permanently and the set-point isn't maintained, calibrate and adjust the controller.
  • B
    Beverly SimmonsAug 25, 2025
    Why is the set-point temperature not reached after the specified time in my Binder VD 115?
    • E
      Erin WilsonAug 25, 2025
      If the set-point temperature isn't reached in your Binder oven after the specified time, ensure the chamber door is completely closed. If the issue persists, the door gasket may be defective and require replacement. Also, calibrate and adjust the controller.
  • H
    Heidi StephensAug 23, 2025
    What to do if my Binder VD 115 Oven chamber doesn’t heat up and shows an “Overtemperature” alarm?
    • J
      Jennifer WarrenAug 23, 2025
      If your Binder oven displays an “Overtemperature” alarm and the chamber doesn’t heat up, allow the chamber to cool down. Acknowledge the alarm on the controller and restart the chamber. If the error occurs again, turn off the chamber.

Summary

1. Safety

4. Installation and Connections

4.5 Electrical connection

Details the electrical connection requirements, including plug types, voltage, and fuses.

5. Functional Overview and Menu Structure of the Controller

6. Start Up

6.4 Vacuum system

Details the operation and connection of the vacuum system for the drying oven.

7. Set-Point Entry

7.1 RD4 controller

Procedure for entering temperature and pressure set-points using the RD4 controller.

7.2 MB2 controller

Procedure for entering set-points via the MB2 controller's menus or normal display.

8. Setting Special Controller Functions

8.5 Activating; deactivating temperature control

Instructions for enabling or disabling the temperature control function.

8.6 Activating; deactivating pressure control

Instructions for enabling or disabling the pressure control function.

8.7 Drying monitoring

Details the process of monitoring the drying cycle based on pressure changes.

9. Authorization Levels and Password Protection

9.2 MB2 controller

Details user management, authorization levels, and password protection for MB2 controller.

9.2.2 Log in

Step-by-step guide for logging into the controller with user credentials.

11. Temperature Safety Devices

11.2 Overtemperature safety controller class 2

Describes the electronic overtemperature safety controller and its protection functions.

11.2.2 Setting the safety controller

Instructions for setting the safety controller mode and value for temperature protection.

12. Tolerance Range Settings

12.1 RD4 controller

Setting tolerance ranges and alarm delay times for temperature and pressure on RD4 controller.

13. Notification and Alarm Functions

13.2 Alarm messages

Details the various alarm messages generated by the controller for equipment failures.

14. MB2 Controller: Timer Program (Stopwatch Function)

15. MB2 Controller: Time Programs

15.4 Creating a new time program

Steps for creating a new time-based program with set-points and durations.

15.5 Program editor: program management

Overview of program management functions like editing, copying, and deleting programs.

16. MB2 Controller: Week Programs

16.3 Creating a new week program

Guide for creating new weekly programs with time-based set-points.

16.4 Program editor: program management

Functions for managing week programs, including editing, copying, and deleting.

18. RD4 Controller: Data Recorder

19. USB Menu: Data Transfer via USB Interface

20. MB2 Controller: Chart Recorder Display

23. Cleaning and Decontamination

24. Maintenance and Service, Troubleshooting, Repair, Testing

24.2 Simple troubleshooting

Provides basic troubleshooting steps for common chamber issues.

25. Disposal

Binder VD 115 Specifications

General IconGeneral
Power Supply230 V, 50/60 Hz
Control TypeDigital
Number of Shelves (standard/max)2/5
Nominal Power2.5 kW
Temperature Range5 °C above ambient temperature to 300 °C
Interior Volume115 L
Power Consumption2.5 kW

Related product manuals