Issue 03/2020 Art. No. 7001-0384
Operating Manual
Translation of the original operating manual
VDL (E3.1)
Vacuum Drying Oven
Standard: with microprocessor program controller MB2
Optional: with microprocessor program controller RD4
Model Model version Art. No.
VDL 23 VDL023-230V 9630-0009
VDL 23-UL VDL023UL-120V 9630-0013
VDL 56 VDL053-230V 9630-0010
VDL 56-UL VDL053UL-120V 9630-0014
VDL 115 VDL115-230V 9630-0011
VDL 115-UL VDL115UL-120V 9630-0015
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder vdl 23

Questions and Answers

  • C
    Claire SmithSep 15, 2025
    Why are there deviations from the temperature set-point in equilibrated state in my Binder vdl 23 Oven?
    • J
      Jeffrey CochranSep 15, 2025
      If you experience deviations from the temperature set-point in equilibrated state in your Binder Oven, use only the delivered expansion racks and do NOT change between aluminum and stainless steel racks.
  • S
    Samantha CollierAug 25, 2025
    Why doesn't my Binder vdl 23 heat up and shows a 'safety controller' alarm?
    • S
      Stacy EllisAug 25, 2025
      If your Binder oven isn't heating up and displays a 'safety controller' alarm, it means the inner chamber temperature has reached the safety controller value, causing the safety controller to respond. Allow the chamber to cool down. Acknowledge the alarm on the controller. Then, check the temperature setpoint and safety controller value settings. If necessary, choose a suitable safety controller value.
  • J
    jessicagarciaAug 25, 2025
    How to fix if the Binder vdl 23 set-point temperature is not reached after specified time?
    • J
      Jeffrey BowersAug 26, 2025
      If the set-point temperature in your Binder oven is not reached after the specified time, there might be a few reasons: * The chamber door may not be properly closed. Ensure it is completely closed. * The door gasket may be defective. Consider replacing it. * The controller may not be adjusted. Calibrate and adjust the controller.
  • T
    Tammy DavisAug 26, 2025
    What to do if my Binder vdl 23 chamber doesn't heat up?
    • R
      randyclarkAug 27, 2025
      If your Binder oven chamber isn't heating up, consider the following: * The pressure threshold of 100 mbar might not be reached, preventing heater release. Wait until the pressure threshold is reached. * The chamber door might not be properly closed. Ensure it is completely closed. * The door gasket might be defective. Replace it. * The Pt 100 sensor might be defective. Contact BINDER service.
  • M
    Madison BarrettSep 16, 2025
    What to do if my Binder vdl 23 Oven has no chamber function (dark display)?
    • P
      Priscilla FlynnSep 16, 2025
      If there is no chamber function (dark display) on your Binder Oven, deactivate standby mode.
  • S
    shawn25Sep 23, 2025
    How to fix a Binder Dryer chamber that is permanently turned off?
    • J
      John WeissSep 23, 2025
      To address a Binder Dryer chamber that is permanently turned off: - Check the connection to the power supply. - Verify the voltage is correct (115V or 230V). - Ensure the oven is connected to the compressed air supply. - Inspect the chamber fuse. - If the safety device has shut off the oven, let it cool, press the reset button, and verify the temperature set-point and safety device settings, choosing an appropriate limit if necessary. - If the safety device is faulty, contact BINDER Service.
  • M
    Mark DavisSep 7, 2025
    Why is my Binder Oven chamber not heating up?
    • C
      csalasSep 8, 2025
      If your Binder Oven chamber isn't heating up, it could be due to several reasons. First, the pressure threshold of 100 mbar might not be reached, preventing heater release, so wait until it is reached. Second, the chamber door might not be properly closed, so ensure it is completely closed. Third, the door gasket could be defective, requiring replacement. If none of these resolves the issue, contact BINDER service because the Pt 100 sensor could be defective.
  • K
    Kevin MillerSep 8, 2025
    What to do if my Binder vdl 23 chamber is heating permanently but not maintaining the set-point?
    • D
      davidleonSep 9, 2025
      If your Binder Oven chamber is heating permanently and the set-point is not maintained, calibrate and adjust the controller. For other potential causes, contact BINDER service.
  • A
    Annette PerezSep 2, 2025
    What to do if acknowledging the alarm does not cancel the alarm state in my Binder oven?
    • M
      Miranda SmithSep 2, 2025
      If acknowledging the alarm does not cancel the alarm state in your Binder oven, it indicates that the cause of the alarm persists. Remove the cause of the alarm. If the alarm state continues, contact BINDER service.
  • M
    Matthew MckinneyAug 31, 2025
    Why are there deviations from the indicated heating-up times in my Binder vdl 23?
    • D
      Dr. Michelle MartinezAug 31, 2025
      If you're experiencing deviations from the indicated heating-up times in your Binder oven, it might be because the oven is fully loaded. Try charging the oven with less material or consider allowing longer heating-up times.