EasyManua.ls Logo

Binder vdl 23 - Connection of an Extraction System at the Pump Module; Vacuum Connection; Figure 18: Position of the Vacuum Connection on the Chamber Rear (Example Size 115)

Binder vdl 23
235 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VDL (E3) 09/2019 Page 63/234
6.3.3 Connection of an extraction system at the pump module
Avoid the solvent accumulation in the pump module as this would cause the pump module to
become an occasionally or continuously / for long periods / frequently potentially explosive
area (Zone 0 or 1). The VDL vacuum drying oven located on top of the module is constructed
in device category 3 in regards to its surroundings.
DANGER
Explosion hazard due to the spread of an explosive atmosphere to unprotected
chamber parts and ignition due to electric sparking or hot surfaces
Serious injury or death from burns and / or explosion pressure.
Provide active extraction (technical ventilation according to country-specific regulations
TRBS 2152 Part 2 for Germany) prior to commissioning the chamber.
Extraction must include the entire installation area of the vacuum drying oven and pump
module; in addition, extraction must be provided on the floor behind the chamber.
Connect an extraction system to the provided port on the rear of the pump module.
6.4 Vacuum connection
Connect the vacuum drying oven to a vacuum pumpor to a domestic vacuum system. For this purpose,
the vacuum connection (6) with small flange DN16 must be connected to the back of the chamber at the
top with the vacuum pump or domestic vacuum system via a vacuum suction hose or a fixed vacuum
pipe.
(6)
Figure 18: Position of the Vacuum connection on the chamber rear (example size 115)
When using a vacuum suction hose BINDER recommends the connection kit for the VP4 pump from
BINDER (chap. 6.4.5). The optional pump module (chap.) provides an appropriate hose port on the rear.

Table of Contents

Other manuals for Binder vdl 23