Section 3
Installation
3.1 Environmental Requirements
To insure correct operation and stable performance over an extended period of time, install
the SmartSpec 3000 in a location which meets the following conditions:
Room temperature between 20 and 35 °C. Not recommended for cold room use (4 °C).
Not exposed to direct sun light.
Not subject to direct or continuous vibration.
Not subject to intense magnetic or electromagnetic fields.
Relative humidity between 0–95%.
Area free from corrosive gases or other corrosive substances.
Area with very little dust or other airborne particles.
Allow a 10 cm minimum space around the instrument for proper air flow.
3.2 Power On and Initial Display
1. Power up SmartSpec. When you turn on SmartSpec 3000, it will go through a self-
diagnostic routine during which it tests the lamp, optical components, diode array, power
supply and other components. If a problem is detected, SmartSpec 3000 will alert you to call
Bio-Rad Technical Support. If boot up is successful and no errors are detected, the Home
screens will be displayed.
alternating with
4
CHOOSE ANY
SMARTSPEC ASSAY
04:23:11 PM SmartSpec
Fri Apr 23, 1999