Outlines systematic approach for diagnosing and fixing failures, emphasizing step-by-step problem-solving.
Mentions a diagram for understanding the equipment's physical layout.
Details causes, problems, and solutions for leakage alarms, advising checks on pipelines and seals.
Addresses heating tube failures, parameter settings, and altitude effects, offering adjustment methods.
Explains air leakage detection and resolution, including checks on fittings and environmental factors.
Covers sensor connection/damage issues and low temperature effects, suggesting sensor or board checks.
Focuses on temperature sensor failures on the circuit board and recommends board replacement.
Describes issues related to insufficient water, pointing to float ball problems and suggesting water level checks.
Explains water shortage causing heating coil issues, involving float, solenoid valve, or overload checks.
Details causes like faulty door switches or loose door shafts, providing solutions for door interlock errors.
Covers daily/weekly cleaning, periodic inspections, and essential safety notes before maintenance procedures.
Details steps for checking and replacing the safety valve, including operational procedures and safety advice.
Provides step-by-step instructions for replacing the thermostat, covering wiring and testing.
Outlines the process for replacing heating tubes, focusing on disconnection, removal, and reinstallation.
Guides on replacing water level floats and temperature sensors, including wiring and testing procedures.
Visual representation of the sterilizer's electrical connections and components.
Illustrates the pipeline system and provides a list of components with their designations.
This document provides a comprehensive maintenance manual for the BIOBASE Vertical Autoclave Sterilizer, BKQ-B Series, model number XDGF2017-010. The manual is issued by BIOBASE BIODUSTRY (SHANDONG) CO., LTD, located at NO. 51 SOUTH GONGYE ROAD, JINAN, SHANDONG, CHINA.
The BIOBASE Vertical Autoclave Sterilizer is designed for sterilization applications, utilizing steam to achieve high-temperature sterilization. Its primary function is to sterilize various items, ensuring they are free from microorganisms. The device is equipped with a control panel for operation and display of status, including temperature and pressure.
The sterilizer operates through a series of steps, including heating, sterilization, and cooling. It features a door detection switch to ensure the door is properly closed during operation, preventing accidental opening while the machine is running. The device includes a water level float to monitor the water level within the pot, which is crucial for the heating process and preventing dry heating of the heating tube. Users can adjust parameters such as the ending temperature of exhausting cold air, which is set to between 101-103 degrees Celsius by default, to optimize sterilization based on specific needs and environmental conditions. The manual provides instructions for accessing and modifying these parameters through the control panel. The device also incorporates safety features such as a pressure gauge, a relief valve, and a safety door interlock to ensure safe operation.
While specific numerical technical specifications are not explicitly listed, the manual implies several key aspects:
The manual emphasizes a systematic approach to inspection and repair, starting with identifying the major fault and then progressively narrowing down the cause. It advises operators to gather information about the machine's usage history, previous maintenance, and specific symptoms before attempting repairs. A critical aspect of maintenance involves reviewing maintenance records to track replaced parts and consumables.
The manual provides detailed solutions for various alarm codes (Err 001, Err 002, Err 004, Err 008, Err 016, Err 032, Err 064, Err 128), which indicate specific malfunctions:
The manual outlines steps for replacing key electrical components, emphasizing the importance of cutting off power and ensuring no pressure inside the sterilizer before any operation. These procedures are intended for qualified personnel.
The manual also includes electrical wiring diagrams (Appendix 1) and pipeline wiring diagrams (Appendix 2) to assist with troubleshooting and maintenance, providing a visual guide to the internal components and their connections. Contact information for BIOBASE GROUP is provided for inquiries, complaints, and after-sales service.