ISOKINETIC PROFILES DEFINED (Clinical Pro Only)
Minimum time is six minutes for all profiles, target resistance level LED will flash during time
duration for that segment. All segments following warm up are based on the actual work rate
performed in the immediate previous segment.
Constant Resistance (Manual Mode)
Speed is consistent for the duration of this profile. The user selects the speed and simply begins
to pedal. Resistance level is how hard the patient pushes against the pedal.
Appropriate for general aerobic endurance
Exercise at lower work rates for longer period of time if de-conditioned or weight loss
is the goal
Constant Level with Warm up and Cool Down
The starting segment (1, warm-up) is however hard the patient chooses to work. The middle
segment (2, work) target value equals 100% more than the average actual work rate performed
in the warm-up segment (rounded to the nearest 10 WATTs). Target resistance level flashes in
the resistance bar during the time duration for the work segment. The cool down segment
equals 50% less than work performed in the work segment.
Includes a warm-up and cool down period
Appropriate for general aerobic endurance
Exercise at lower work rates for longer period of time if de-conditioned or weight loss
is the goal