15. DIAGNOSTICS
DIAGNOSTICS — 34 —
There are two hidden key presses that will allow for either configuration, setup or
hardware debug.
Configuration Mode
To enter the configuration setup mode, power the unit ON while holding the select profile
button down during power on LED test. Use the +/- adjust buttons to toggle through the
selections. The selections for configuration setup are as follows:
ENGL/EURO Description
1 UBC/ BIO UBC or BioStep
CPRO or STD For Clinical Pro BioStep with both
HRS Constant Torque and Isokinetic resistance or
Standard BioStep with constant torque only.
Diagnostic Mode
(For 950-214 and 950-220. See NOTE at end of this section for 950-215.)
To enter the debug mode, power the unit ON holding the select exercise button down during
power on LED test. Use the select exercise button to cycle through the selections.
The selections for hardware debug mode are as follows:
Description
1. Load PWM / Speed display Ramps PWM value from 0-255 (0-100%) displayed in upper
display window (use <L> to increase resistance)..
2. Batt Displays battery voltage in lower window
NOTE: The <Select Exericse> button is hidden but functional on model 950-215. To activate, press down
on the display 2-3/4” below the <Select Profile> button while holding down <M>.