EasyManua.ls Logo

Biolase iLase - 6 Cleaning and Sterilization; Cleaning and Disinfection

Biolase iLase
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5400230 Rev. H
19
6 Cleaning and Sterilization
6.1 CLEANING AND DISINFECTION
CAUTION:
During use the internal components of the iLase under the cover may become hot to the touch.
Allow the iLase to cool for a few minutes before removing the cover for cleaning and sterilization.
The iLase must be cleaned prior to use on each patient.
Cleaning must be performed within a
maximum of one (1) hour after the procedure, and prior to sterilization.
Manual cleaning is
recommended.
Prepare a cleaning solution per the manufacturer’s instructions. Use a commercially available
surgical instrument detergent/enzymatic cleaning solution with a pH of 7.0, such as Enzol®, or a
similar enzymatic presoak and cleaner. Follow the instructions for disposal of used solution.
To disinfect the iLase main body and battery, perform the following steps in the indicated order:
1
Place the iLase in STANDBY mode.
2
Wear protective gloves and eyewear when handling the contaminated main body and battery.
3
Remove the used tip from the cover (if present) and dispose of in a biohazard medical waste
Sharps container.
4
Remove the battery from the main body and set it aside.
5
Pull the iLase cover forward and slide it off the body of the main body.
6
To disinfect the main body, wipe with an appropriate disinfecting solution,such as Cavidice
®
or a
similar quaternary ammonium compound product (containing 20% alcohol or less) and follow the
manufacturer’s instructions.
CLEANING THE COVER
1
Rinse the cover under lukewarm water (22 - 43º C) for a
minimum of 10 seconds
to remove gross soil.
2
Soak a piece of gauze in the cleaning solution and wrap the handpiece cover. Leave it
wrapped for a minimum of 10 minutes. Unwrap, and using a soft-bristled brush dipped in the
cleaning solution, gently brush the cover for
at least 15 seconds
.
3
Rinse the cover under lukewarm running tap water (22 - 43º C) for a minimum of 10 seconds and
dry with a lint-free cloth.
4
Visually inspect the cover for any residual soil. If present, repeat steps 1 through 3 until residual
soil is removed.