ABP-01
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5% UT (>95%
dip) for 0.5 cycle;
40% UT (60% dip)
for 5 cycles; 70%
UT (30% dip) for
25 cycles; <5% UT
(>95% dip) for 5
sec. Note: UT is the
nominal voltage of
mains.
Not
applicable
Mains power quality is that of
a typical commercial and/or
hospital environment.
ABP-01
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic
fields are at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
ABP-01
Conducted RF IEC
6100-4-6
3V eff 150KHz-
80MHz
Not
applicable
Mains power quality is that of
a typical commercial and/or
hospital environment
ABP-01
Radiated RF IEC
61000-4-3
3V/m 80MHz-
2,5GHz
3V/m
Recommended Separation
distance: d=[3,5/V1]P
d=[3,5/3V/m]P (80MHz
– 800MHz) d=[7/3V/m]P
(800MHz – 2,5GHz) where:
P is the highest radiated
power disclosed by the
manufacturer of transmitter
[W] and d is the recommended
separation distance [m].
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey1 a should be less than
the compliance level in each
frequency range 2 You may
observe disturbance nearby
any of those equipment which
has the following indication:
Note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is applicable.
Note: these are guidelines. Actual conditions may vary.
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with