1. With the blood pressure monitor off, press and hold the button until appears on the screen,
release the button.
2. While is flashing on the screen press the button again to clear all the values. You will hear a
beep and the screen will then go back to displaying the time.
4. Error Messages / Malfunctions
If an error occurs during a measurement the LCD displays the corresponding error code.
Error Possible Cause Remedy
Signal too weak
The pulse signals on the cuff are too weak. Re-position
the cuff and repeat the measurement.
Error signal
During the measurement, error signals were detected
by the cuff, caused for instance by movement or muscle
tension. Repeat the measurement, keeping your arm still.
No pressure in the cuff
An adequate pressure cannot be generated in the cuff. A
leak may have occurred. Check that the cuff is correctly
connected and is not too loose. Replace the batteries if
necessary. Repeat the measurement.
Abnormal result
The measuring signals are inaccurate and no result can
therefore be displayed. Read through the checklist for
performing reliable measurements and then repeat the
measurement.
Pulse or cuff pressure
too high
The pressure in the cuff is too high (over 300 mmHg) OR
the pulse is too high (over 200 beats per minute). Relax
for 5 minutes and repeat the measurement.
Pulse too low
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
If problems occur when using the device the following points should be checked, and if necessary, the
corresponding measures should be taken.
Malfunction Remedy
The display remains blank when
the device is switched on.
1. Check batteries for correct polarity.
2. If using the AC adapter, ensure it is correctly inserted.
The pressure does not rise even
though the pump is running.
Check the connection of the cuff tube and connect properly if
necessary.