NOTES: { }-command data, [ ] -return data, [OK] = "OK", [CR] = 0dh ???? = 30h, 30h, 30h, 30h -39h, 39h, 39h, 39h
(4 bytes data) ??? = 30h, 30h, 30h -39h, 39h, 39h (3 bytes data) ?? = 30h, 30h – 39h, 39h (2 bytes data) #### = 30h,
30h, 30h, 30h – 3Fh, 3Fh, 3Fh, 3Fh (4 bytes data) ### = 30h, 30h, 30h – 3Fh, 3Fh, 3Fh (3 bytes data) ## = 30h, 30h
– 3Fh, 3Fh (2 bytes data) # = 30h –3Fh (1 byte data) <address> 30h, 30h -3fh, 3fh (2 bytes data)
CURR <address> current {001-XXX} <CR> [OK] [CR]
Set Current Level
SOVP <address> voltage {010-XXX} <CR> [OK] [CR]
Set Upper Voltage Limit of
PS
SOUT <address> 0 <CR> [OK] [CR]
Disable Output of PS
SOUT <address> 1 <CR> [OK] [CR]
Enable Output of PS
PROM <address> location {1-9} Voltage {010-XXX} Current {001XXX}
<CR> [OK] [CR]
Set Voltage and Current
values of Preset Memory
PROP <address> location {00-19} Voltage {010-XXX} Current {001XXX}
Minute {00-99} Second {00-59} <CR> [OK] [CR]
Set Voltage, Current and
Time period of Timed
Program
RUNM <address> location {1-9} <CR> [OK] [CR]
Recall Preset Memory 1-9
RUNP <address> times {0000-9999} <CR> [OK] [CR]
Run Timed Program (0000
= run infinite times)
STOP <address> <CR> [OK] [CR]
Stop Timed Program