Mounting and commissioning
20 by Blum-Novotest 2012
3.4 Mounting of the stylus
CAUTION
Damage of the measuring mechanism
• TC60: When mounting the stylus into the probe, it is absolutely
necessary to lock the probe with the mounting tool (B) to avoid
torsion of the measuring mechanism.
• TC63-30 / TC76:
When mounting the stylus into the probe, it is absolutely necessary
to use the interlocking disk (A) (backlash-free).
Pls. take care of the orientation of the interlocking disk (A)
(shark360° = top)
• Remove interlocking disk before measurement
• Calibrate probe after mounting of stylus!
• Damaged styli are not reusable!