BPAP Series User Manual V1.0
40 / 58
When the airflow is on, an audible alarm
will sound if the airway pressure is below
the warning limen; the screen will display
―Low Pressure!!‖.
Note: The limens for different models:
Off,3 ~ 24 hPa applies to G3 B25VT, in
0.5 hPa increments, the default setting is
―4 hPa‖.
Off, 3 ~ 29 hPa applies to G3 B30VT, G3
B30SV and G3 LAB, in 0.5 hPa
increments, the default setting is ―4 hPa‖
Low RR!!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
When the airflow is on, an audible alarm
will sound if the respiratory rate is below
the limen; the screen will display ―Low
RR!!!‖.
Setting range: Off, 4 ~ 40 BPM, in 1
BPM increments, the default setting is ―6
BPM‖.
Note: This function is available under the
work mode of S/T or T
When SpO
2
Kit is applied, an audible
alarmwill sound when the value of SpO
2
is
lower than the warning threshold; the
screen will display ―Low SpO
2
!!!‖.
Setting range: Off, 70% ~ 100%, in
1% increments, the default setting is
―85%‖.
Note: This function is available only
when the device is equipped with SpO
2
Kit
When the airflow is on, an audible alarm
will sound if the air leak rate exceeds 150
L/min; the screen will display ―Leak!!‖.
The alarming duration time is no less than
30s.
Mask Blocked!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
When the airflow is on, an audible alarm
will sound if the vents of the mask are
blocked; the screen will display ―Mask
Blocked!!‖
Low MV!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
When the airflow is on, an audible alarm
will sound if the minute volume is below
the warning limen; the screen will display
―Low MV!!‖.
Setting range: Off, 1 ~ 30 L/min, in 1
L/min increments, the default setting is