BPAP Series User Manual V1.0
41 / 58
If the voltage supplied by power adaptor
is lower than 22V, an audible alarm will
sound and the screen will display ―Low
Input Voltage!!‖
High RR!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
When the airflow is on, an audible alarm
will sound if the respiratory rate exceeds
the threshold; the screen will display
―High RR!!‖.
Setting range: Off, the setting value of
Low RR ~ 80 BPM, in 1 BPM increments,
the default setting is ―40 BPM‖.
Note: This function is avaliable under the
work mode of S/T or T
When humidifier is applied, an audible
alarm will sound when the humidifier fails
to work in 10 minutes; the screen will
display ―Humidifier Failure!!‖
When the Filter Alarm feature is enabled,
an audible alarm will sound if the preset
replacement time reaches but without
replacing the air filter; the screen will
display ―Please Change Filter!‖. The
default setting is ―Off‖.
When the tubing Alarm feature is
enabled, an audible alarm will sound if
the preset replacement time reaches but
without replacing the tubing; the screen
will display ―Please Change Tube!‖
When the Mask Alarm feature is enabled,
an audible alarm will sound if the preset
replacement time reaches but without
replacing the mask; the screen will
display ―Please Change Mask!‖
The screen will display ―SD Card Full!‖ if
the SD card has reached its maximum
capacity
The screen will display ―Reinsert SD
card!‖ if the SD card fails to work
Note: the delay time of alarming system of this device is no more than 1s.