Recommended separation distances between RF wireless communications
equipment
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between RF wireless
communications equipment and the device as recommended below, according to the
maximum output power of the communications equipment.
Frequency
MHz
Maximum
Power W
Distance
IEC 60601 Test
Level
Compliance Level
385
1.8
0.3
27
27
450
2
0.3
28
28
710
0.2
0.3
9
9
745
780
810
2
0.3
28
28
870
930
1720
2
0.3
28
28
1845
1970
2450
2
0.3
28
28
5240
0.2
0.3
9
9
5500
5785
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC M1 Mini and is the answer not in the manual?

BMC M1 Mini Specifications

General IconGeneral
CategoryMedical Equipment
Pressure Range4-20 cmH2O
TypeCPAP
Weight0.8 kg
HumidifierIntegrated
Power SourceAC power adapter, optional battery pack