Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
7 / 49
Static Pressure Stability at 10 cmH
2
O
±0.5 cmH
2
O
Dynamic Pressure Stability
Pressures (cmH
2
O)
10 BPM
15 BPM
6.5
±0.5
±0.5
10
±1
±1
20
±1
±1
Device with humidification and 22 mm Tubing or Heated Tubing.
Ramp
The ramp time ranges from 0 to 60 minutes.
The A-weighted sound pressure level and sound power level
When the device is working at the pressure of 10 hPa, its sound pressure level and
sound power level shall not be greater than the values in the following table:
Sound Pressure
Level
Uncertainty
Sound Power
Level
Uncertainty
28 dB(A)
2 dB(A)
36 dB(A)
2 dB(A)
Note: Declared dual-number noise emission values in accordance with ISO
4871:1996.
Maximum Flow
Test Pressures (cmH
2
O)
4
8
12
16
20
Measured Pressure at the Patient
Connection Port (cmH
2
O)
3
7
11
15
19
Average Flow at the Patient
Connection Port (L/min)
85
135
140
140
140
When the working pressure is set to the values listed in the table, the average
flow rate at the patient end should be greater than 80% of the corresponding
flow value in the table.
Air Tubing
Air tubing
Length
Inner diameter
Tubing
6 ft. (1.83 m)
19 mm
Heated Tubing
6 ft. (1.83 m)
19 mm
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
Air filter
Type
Material
Average arrestance
Standard filter
Polyurethane
> 20% for 10 micron

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC REACTHEALTH Luna G3 APAP and is the answer not in the manual?

BMC REACTHEALTH Luna G3 APAP Specifications

General IconGeneral
BrandBMC
ModelREACTHEALTH Luna G3 APAP
CategoryRespiratory Product
LanguageEnglish

Related product manuals