Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
33 / 49
18.1 Cleaning the Water Chamber
(1) Opening the Water Chamber: Open the cap of the water chamber, as shown in
Fig. 18-1.
Fig. 18-1
(2) Cleaning the Water Chamber: Dilute Alconox to 1% with water 45°C to 60°C
(113°F to 140°F). Soak the water chamber in the detergent for 5 minutes. Clean the
water chamber with a soft bristled brush for one minute. Pay particular attention to
all gaps and cavities. Then rinse the water chamber with running water for 5 minutes.
Wipe it dry with a soft cloth or air dry it away from direct sunlight.
(3) Putting the Water Chamber back: according to instructions in14.1.3.
WARNINGS!
• Emptying and cleaning the water chamber daily will help prevent mold and
bacteria growth.
• Allow the water in the chamber to cool down to room temperature before
removing it from the device.
CAUTIONS!
• Clean the water chamber only after the water in it cools. Make sure that no water
enters the device.
• After cleaning, rinse the water chamber thoroughly in clean water to make sure
that no soap residue is left; then wipe it dry with a lint-free cloth, so as to prevent
calcareous accumulation.
• Check the water chamber for any leak or damage. Replace the water chamber
if there is any damage.
• It is recommended to clean the water chamber and change the water daily.
18.2 Cleaning the Transfer Box
(1) Removing the Transfer Box: First remove the water chamber from the device,
and then remove the transfer box, as shown in Fig. 18-2.