17. Alert
An audible alert will sound in 6s if the device is accidentally
disconnected from power when it is delivering air.
Note:
(1) The alert will not sound if power failure occurs when the
device is in standby state.
(2) No alert message on the screen during a power failure.
An audible alert will sound if no airflow comes out of the
machine; the screen will display “Device Fault!!!”.
When the airflow is on, an audible alert will sound if the air
leak rate is excessive; the screen will display “Leak!!”.
If the voltage supplied by power adaptor is lower than 22 V,
an audible alert will sound and the screen will display “Low
Input Voltage!!”.
When humidifier is applied, an audible alert will sound when
the humidifier fails to work; the screen will display “Humidifier
Failure!!”.
When the Filter reminder feature is enabled, an audible alert
will sound if the preset replacement time reaches but without
replacing the air filter; the screen will display “Please Change
Filter!”.
When the tubing reminder feature is enabled, an audible
alert will sound if the preset replacement time reaches but
without replacing the tubing; the screen will display “Please
Replace Tubing!”.
When the Mask reminder feature is enabled, an audible alert
will sound if the preset replacement time reaches but without
replacing the mask; the screen will display “Please Replace
Mask!”.
The screen will display “SD Card Full!” if the SD card has
reached its maximum capacity.
The screen will display “Reinsert SD Card!” if the SD card fails
to work.